Educerie · AP · Chemistry
AP Chemistry 3.13
Beer-Lambert Law
How can you find the concentration of a chemical species in a mixture?
What you will learn
- Understand how absorbance depends on molar absorptivity, path length and concentration
- Know that absorbance is proportional to concentration when wavelength and path length are fixed
- Understand why the wavelength of maximum absorbance is used
- Know how a calibration curve gives an unknown concentration
You must be able to
- Calculate absorbance, concentration, path length or molar absorptivity with units
- Predict how absorbance changes when concentration or path length changes
- Read an unknown's concentration from a calibration line and undo any dilution
- Explain why the wavelength of maximum absorbance gives the best sensitivity
- Name an error source and the direction it moves the result
- Describe how to prepare and use standards with a blank
In the exam. Section I multiple choice: a calculation with , a prediction of how absorbance changes when concentration or path length changes, a reading from a calibration line, or a question on what could make a measurement wrong. Section II: a spectrophotometry short question appeared in 2026, and similar parts sit inside longer experiments. Marks go to the set-up with units, to using the calibration line, and to a reason that mentions the number of absorbing particles in the path.
1The equation
What each term means:
- , absorbance. A number with no unit. A higher value means more light absorbed. Zero means none absorbed.
- , molar absorptivity. How strongly one mole of this species absorbs light of this wavelength. Its unit is . A species that absorbs intensely has a large . It depends on the species and the wavelength.
- , path length. The distance the light travels through the solution. In a standard cuvette it is .
- , concentration. The molarity of the absorbing species, in .
Check the units: leaves no unit. That is why has none.
Why is it proportional to and ? The absorbing particles in the beam are what absorb. A longer path crosses more particles. A higher concentration has more particles in every centimetre. So and both count the particles in the light's path: the CED says that and are proportional to the number of light-absorbing particles in the light path.
2Worked examples
Example 1: find . A solution in a cuvette has at a wavelength where .
Example 2: find . A solution has in a cuvette.
Example 3: changes without numbers.
| Change | Effect on |
|---|---|
| Concentration doubled | doubles |
| Concentration halved | halves |
| Path length doubled (a cuvette) | doubles |
| A different species with a larger at the same and | is larger |
| The solution is diluted from to | falls to a quarter |
Example 4: path length. A solution has in a cuvette with . Then . Using by mistake would give , double the true value.
3Absorbance and transmitted light
The instrument actually measures the fraction of light that passes through the solution, the transmittance, . It converts this to absorbance. The conversion is a logarithm:
Landmarks:
Absorbance and transmittance run in opposite directions. As the absorbance rises, the transmittance falls. An absorbance of means that only of the light gets through. The equation sheet gives only , so questions give the absorbance.
4Choosing the wavelength
A species absorbs different wavelengths to different extents. A plot of absorbance against wavelength has peaks. For a measurement of concentration, the spectrophotometer is set to the wavelength of maximum absorbance, the optimum wavelength.
Why? At that wavelength is at its largest. A small change in concentration then gives the largest change in absorbance, so the measurement is the most sensitive. The line on a calibration curve is steepest, and small differences are easiest to tell apart.
Two points to hold on to:
- Fix the wavelength for the whole experiment. changes with wavelength, so mixing wavelengths breaks the proportionality.
- Fix the path length. Use the same cuvette type for every reading.
The colour of a solution is the light it does not absorb. A solution that looks blue absorbs mostly orange and red light, so the instrument should be set in that range.
5The calibration curve
To find an unknown concentration without knowing :
- Prepare solutions of known concentrations (standards) by dilution from a stock solution.
- Set the spectrophotometer to the wavelength of maximum absorbance, and zero it with a blank: a cuvette of solvent only, or the solution without the absorbing species.
- Measure the absorbance of each standard in the same cuvette.
- Plot absorbance against concentration. The points fall on a straight line through the origin.
- Measure the unknown's absorbance, find it on the line, and read across to the concentration.
Figure 1 shows four standards and the unknown. Read it in this order:
- The dots are standards at , , and in units of , with absorbances , , and .
- The line passes through the origin. With , its slope is .
- The unknown has . Reading across to the line and down gives . It lies between the and standards, as it should.
Use only values that lie within the range of the standards. Reading far beyond the highest standard assumes that the line continues, and you have no evidence for that.
A dilution before measuring. If the unknown was too concentrated, it is diluted first. Suppose of the original is diluted to and then reads . The diluted solution is , and the original is five times as concentrated.
6Sources of error (2.E)
The exam asks you to name an error and say which way it moves the result.
| Source | Effect on the reading | Effect on the calculated |
|---|---|---|
| Fingerprints or dust on the cuvette | scatters or absorbs light, reads high | too high |
| Bubbles in the beam | reads high | too high |
| Instrument not zeroed with the blank | all readings shifted | the calibration line does not pass through the origin; wrong results |
| Wrong wavelength | lower sensitivity | less precise |
| Different cuvettes with different path lengths | varies for a reason other than | unreliable |
| Standards made inaccurately | the calibration line is wrong | the unknown is wrong |
| Cuvette not rinsed with the next solution | carry-over from the earlier one | wrong |
Very concentrated solutions can also fall off the straight line. If your highest standard looks flat compared to the line, dilute the unknown into the range.
Practice questions
30 questions on this note, from easy to hard. Each one you get right lights a star of this topic's constellation.